This page allows you to edit the calibration order that is already created. Tell me more.
Select the “Edit Calibration Order” link in the “Select Calibration Order” page. Alternatively, select the hyperlinked calibration order in the multiline of the same page.
The “Edit Calibration Order” page appears.
If you wish to modify a calibration order other than the calibration order selected in the previous page, enter the following field:
|
Calib. Order |
The calibration order you wish to modify (Alphanumeric, 18). |
|
Press <Enter>. The system retrieves the details of the calibration order. | |
|
The system displays the following fields: |
|
|
Date |
The date on which the calibration is done on the equipment. |
|
Status |
The status of the calibration order, which can be “Fresh” or “Closed”. |
Modify the following details, if required::
|
Desc. |
The description of the calibration order (Alphanumeric, 150). |
|
Zoom facility available. | |
|
Equipment |
The equipment for which calibration must be done (Alphanumeric, 30) Mandatory. Ensure that the equipment is already defined in the “Create Equipment“ activity of the “Equipment Induction“ business component and is in “Active” status. Also ensure that the equipment is of type “Normal”, “Rotable” or “Tool”. |
|
Help facility available. | |
|
|
The description of the equipment is displayed alongside. |
|
Zoom facility available. | |
|
Task/Proc. |
Use the drop-down list box to specify whether you wish to attach a task or a standard procedure to the calibration order. The system provides the following options: |
| |
| |
|
Code |
The identifier of the task or the standard procedure that you wish to attach to the calibration order (Alphanumeric, 30). If you are specifying a task, ensure that the task is already defined in the “Create Task” activity and is in the “Active” status. If you are specifying a standard procedure, ensure that the procedure is already defined in the “Create Procedure” activity and is in the “Active” status. |
|
Help facility available. | |
|
The description of the task or standard procedure is displayed alongside. | |
|
Zoom facility available. | |
|
Calib. Type |
Use the drop-down list box to specify the calibration type. The system lists all the calibration types that have been defined in the “Define Calibration Types” activity of the “Calibration Schedule” business component, that are in the “Active” status. Leave the field blank if you do not wish to specify the calibration type. |
|
Priority |
Use the drop-down list box to modify the priority assigned to the calibration order. The system lists all the priorities that have been defined in the “Define Priority” activity of the “Maintenance Organization” business component, that are in the “Active” status. Leave the field blank if you do not wish to assign a priority for the calibration order. |
|
Work Group |
The work group that is responsible for executing the calibration order (Alphanumeric, 30 ) Mandatory. Ensure that the work group is already defined in the “Create Work Group” activity of the “Work Group” business component and is in the “Active” status. |
|
Help facility available. | |
|
Scheduled Calib. Date |
The date on which the calibration is scheduled on the equipment (Date Format). |
|
Actual Calib. Date |
The date on which the equipment is actually calibrated (Date Format). |
Note: Data entry in this field is mandatory, if you wish to close the calibration order.
The system displays the following field:
|
Last Calib. Valid Till |
The date till which the last calibration done on the equipment is valid. |
Enter the following fields:
|
Comments |
Any additional information about the calibration order (Alphanumeric, 255). |
|
Zoom facility available. | |
|
Calibrated By |
The login ID of the user who raised the calibration order (Alphanumeric, 10). Ensure that the employee entered in this field is already defined in the ”Employee Information” business component of Ramco Human Resource Management Solution. |
|
Help facility available. |
The system retrieves the parameters for which calibration is done on the equipment, in the “Calib. Details” multiline.
Modify the following fields, if required:
|
Parameter |
The parameter for which the equipment must be checked for calibration (Alphanumeric, 30). Ensure that the parameter is already defined in the “Create Parameter” activity and is in the “Active” status. Also ensure that the parameter is already attached to the equipment in the “Create Equipment” activity. |
|
Help facility available. |
The system displays the following fields:
|
Type |
The type of the parameter, which can be “Attribute”, “Consumption” or “Range”. |
|
UOM |
The unit of measurement of the consumption or range parameter. |
Enter the following fields:
|
Set Point |
The value of the parameter at which the calibration must be done on the equipment (Alphanumeric, 30). Mandatory. Ensure that you enter a numeric value if you have specified a consumption or a range parameter. If you have specified an attribute parameter, ensure that the non-numeric value that you enter already exists in the list of attribute parameter values defined in the “Create Parameter” activity of the “Equipment Induction” business component. |
|
Help facility available. | |
|
Pre Calib. |
The reading on the equipment before calibrating (Alphanumeric, 30). Ensure that you enter a numeric value if you have specified consumption or a range parameter. If you have specified an attribute parameter, ensure that the non-numeric value that you enter already exists in the list of attribute parameter values that are already defined in the “Create Parameter” activity of the “Equipment Induction” business component. |
|
Help facility available. | |
|
Post Calib. |
The reading on the equipment after calibrating (Alphanumeric, 30). Ensure that you enter a numeric value if you have specified consumption or a range parameter. If you have specified an attribute parameter, ensure that the non-numeric value that you enter already exists in the list of attribute parameter values that are already defined in the “Create Parameter” activity of the “Equipment Induction” business component. |
|
Help facility available. | |
|
Master Instrument Type |
Use the drop-down list box to select the type of the master equipment used for calibration. The system provides the following options: |
| |
| |
|
Master Instrument |
The equipment to be used as the master equipment for calibration (Alphanumeric, 30). If the “Master Instrument” is of type “Internal”, ensure that the equipment entered in this field
|
|
Help facility available. | |
|
UL Of Tolerance |
The upper limit of the tolerance range that can be allowed for the equipment to be calibrated (Alphanumeric, 30). Ensure the following: |
| |
| |
| |
|
Help facility available. | |
|
LL Of Tolerance |
The lower limit of the tolerance range that can be allowed for the equipment to be calibrated (Alphanumeric, 30). Ensure the following: |
| |
| |
| |
|
|
Help facility available. |
|
Calib. Date |
The date on which the calibration is done (Date Format). |
|
Calibrated By |
The login ID of the user who calibrated the equipment (Alphanumeric, 10). Ensure that the employee entered in this field is already defined in the “Employee Information” business component of Ramco Human Resources Management Solution. |
The system displays the following field in the multiline:
|
Calib. Result |
The result of the calibration, which can be “Pass” or “Fail”. The system displays “Pass”, if the value displayed in the “Post Calibration” field falls within the range displayed between the values displayed in the “UL of Tolerance” and the “LL of Tolerance” fields. If the value falls beyond this range, the system displays “Fail”. |
Enter the following field in the multiline:
|
Remarks |
Any additional information regarding the calibration of the equipment (Alphanumeric, 255). |
Enter the following fields:
|
Calib. Valid From |
The date, starting from which, the calibration done on the equipment is valid (Date Format). Data entry in this field is mandatory, if you wish to close the calibration order. |
|
Calib. Valid Till |
The date till which the calibration done on the equipment is valid (Date Format). Data entry in this field is mandatory, if you wish to close the calibration order. Also ensure that the date entered in this field is equal to or greater than the date entered in the “Calibration Valid From” field. |
|
Calib. Result |
Use the drop-down list box to set the result of calibration as “Pass” or “Fail”. The system provides the following options: |
| |
|
Click one of the following radio buttons in the “Change Status To” field:
|
Fresh |
Click this radio button if you wish to change the status of the calibration order to “Fresh”. |
Note: You cannot change the status of a “Closed” or “Cancelled” calibration order to “Fresh”.
To close the calibration order
|
Close |
Select this radio button if you wish to close the calibration order. |
Note: You can close the calibration order only if it is in the “Fresh” status.
In the “Option Settings” activity of the “Clearance Permit” business component, if you have specified that closure of calibration orders is not allowed without returning the clearance permits, the system will not allow you to close the calibration order for which clearance permits exist in an “Issued” status.
To cancel the calibration order
|
Cancel |
Select this radio button if you wish to cancel the calibration order. |
Note: You cannot cancel a calibration order that is in “Closed” status.
Click the “Update” pushbutton, to update the modified details of the calibration order.
The system does the following:
The “Calibration Result” field is set to “Pass”, if the value entered in the “Post Calibration” field falls within the values entered in the “LL of Tolerance” and the “Ul of Tolerance” fields.
The “Calibration Result” field is set to “Fail”, if the value entered in the “Post Calibration” field falls beyond the values entered in the “LL of Tolerance” and the “Ul of Tolerance” fields.
If the “Close” option is selected, the status of the calibration order is set to “Closed” and the “Status” field is updated. If the calibration order is raised from a calibration schedule of “interval” type, the system updates the calendar and usage based schedules as follows:
|
|
Calculates the next due date for the calendar based schedule by adding the actual date on which the calibration is done and the interval. |
|
|
Calculates the usage based schedule by adding the current parameter value and the interval. |
Note: The system allows you to close a calibration order only if there are no clearance permits in the “Issued” status, raised for the calibration order.
If the “Cancel” option is selected, the system cancels the calibration order and sets its status to “Cancelled”.
The system displays the following fields in the “Record Statistics” group box:
|
Created By |
The login ID of the user who created the calibration order. |
|
Created Date |
The date and time at which the calibration order is created. |
|
Last Modified By |
The login ID of the user who modified the calibration order. |
|
Last Modified Date |
The date and time at which the calibration order is modified. |
Click the “Calibration Report” pushbutton to generate a report for the calibration order.
To proceed, carry out the following
Select the “Create Clearance Request” link from the bottom of the page to create clearance request for a calibration order that is in the “Fresh” status.
Note: Refer the topic, “Creating clearance request” for more details.
Select the “View Clearance Details” link from the bottom of the page to view the details of clearance permits raised with reference to the calibration order.
Editing calibration order – An overview
This page allows you to edit the calibration order that is already created. You can modify the following details of the calibration order:
the description
the equipment to be calibrated
the task or standard procedure that must be performed for executing the calibration order
the type and priority of the calibration order
the scheduled date of the calibration as well as the actual date of calibration
the user who has calibrated the equipment
the parameters for which the equipment is measured for calibration
the parameter value, at which the equipment is measured
the pre-calibration and post-calibration measurement of the equipment
the master equipment compared to which the equipment is being calibrated
the permissible tolerance range of the measurement of the parameter.
You can also change the validity period of the calibration done on the equipment and also close the calibration order. On closing the calibration order, the system records the date on which the next calibration is due on the equipment.
You can also cancel the calibration order in this activity. The system cancels the order and sets the status of the calibration order to “Cancelled”.
“Create Equipment” activity under the “Equipment Induction” business component.